Who this guideline covers
Neonatal, pediatric, and adult patients with an artificial airway
Read the original guidelineKey recommendations
AARC combined a systematic review with ratings from a modified RAND/UCLA appropriateness process. The evidence level and panel rating describe different things.
Identify an indication before suctioning
Visible secretions, breath-sound changes, and a sawtooth ventilator waveform can support suctioning in adults and children. An abrupt rise in airway resistance can be a clue in neonates.
Level B · panel rating 8 for adult and pediatric signs, median 7 for the neonatal sign
Use assessment to guide timing
The neonatal and pediatric evidence reviewed by AARC supports suctioning when indicated. Outcomes were similar with as-needed and scheduled suctioning in that evidence.
Level B · panel rating 8
Prepare for oxygenation and choose a system
AARC supports preoxygenation for pediatric and adult patients. It found both closed and open systems effective for adult secretion removal. Open suction uses sterile technique.
Preoxygenation B / 9 · system choice B / median 8.3 · sterile technique C / 7
Favor shallow suction and avoid routine saline instillation
Start with a shallow technique and reserve deep suctioning for cases where shallow suction is ineffective. The panel advises generally avoiding saline instillation during suctioning.
Level B · median ratings 7.7 for shallow suction, 7.2 for reserving deep suction, and 8.85 for saline
Limit the duration and vacuum
The panel advises using the least vacuum that effectively clears secretions and keeping the suction event brief, with an upper limit of 15 seconds. Reassess the patient’s response and the indication before repeating.
Level C for duration and vacuum · panel rating 9 for lowest effective vacuum
Evidence ratings refer to the original AARC guideline.
Clinical context
The guideline gives two versions of the sizing cutoff
The abstract and detailed narrative give <70% for neonates and <50% for pediatric and adult patients. Table 1 gives <70% across age groups. Both passages are in the same publication. The PulmTools calculator displays both diameter-ratio comparisons so the reader can see which assumption is being used and compare it with local policy and device labeling.
Diameter ratio and area occupied describe different measurements
Diameter ratio compares catheter outer diameter with airway inner diameter. Estimated area occupied squares that ratio for circular cross sections. For example, 70% of diameter corresponds to approximately 49% of area. When reviewing a cutoff, identify the quantity being compared before applying it to a catheter size.
Practice case
An adult is ventilated through an endotracheal tube. The ventilator waveform develops a sawtooth pattern and secretions are visible in the tube.
Which findings support assessing the need for suction?
Show the reasoning
The waveform change and visible secretions both support an assessment for suctioning. Reassess secretion clearance and tolerance afterward, including oxygenation and changes in the waveform.
Related PulmTools resources
References
- AARC Clinical Practice Guidelines: Artificial Airway Suctioning
Blakeman TC, Scott JB, Yoder MA, Capellari E, Strickland SL. Respir Care. 2022;67(2):258–271. doi:10.4187/respcare.09548.
Primary source. Read the abstract, the catheter-size discussion, and Table 1 together when reviewing sizing.
- Comparison of three suction catheter sizing methods
Russian CJ, Gonzales JF, Henry NR. Respir Care. 2014;59(1):32–38. doi:10.4187/respcare.02168.
Bench and mathematical context for the difference between diameter ratio and estimated area occupied.
For education. Use patient-specific assessment, the full guideline, and local protocols for bedside decisions.